Implant Handpiece Maintenance Process
1. Wipe the surface of the body with alcohol(It is recommended to clean it within 5 minutes after the operation. If it is not cleaned for a long time, blood will coagulate.)
2. Clean spindle(Use the senda oil spray can to press 2-3 times into the spindle, and then press 2-3 times into the hole of back cover, and then wipe it with a cloth.)
3. Internal cleaning
① Manual cleaning(Install spray nozzle on the handpiece, use the SENDA oil spray against the handpiece to press, until the liquid sprayed from the head becomes clean.)
② Machine cleaning (put in lubricating device machine, operate according to the manual.)
4. Blow out liquid inside (After cleaning use the 3-way syringe to blow out excess liquid, prevent oil drenching after high-temperature sterilization.)
5. Disinfection and sterilization (Use sterile bag packing, and put the handpiece in Sterilizer for high temperature sterilization.)
6. Storage (Store the handpiece in a dry place after sterilization.)
7. Handpiece Lubricating (When the handpiece is sterilized, the bearing is very dry, and the bearing is easy to wear when used directly. Dropping two drops of lubricating oil can prolong the life of the handpiece.)